Operating Principle When a magnetic base is brought near to a ‘U’ magnet, leakage magnetic flux gets changed. This change deflects a magnetic needle placed in the flux area. The deflection is calibrated to indicate the air gap (non-magnetic coating) between ‘U’ magnet & magnetic base.
Easy to operate After presetting the instrument place it on the job, press a button and you get the reading. In a short time number of jobs can be checked.
Specifications | Accuracy as % of obtain reading(or +-2.5) | Min. base thickness mm | Min. Area of measurment mm *mm | Min. radius of curvature mm | Dimention mm | Weight gms |
---|---|---|---|---|---|---|
Coatmeter | 10 | 1.5 | 20 x 50 | 50 | H 80 W 60 B 30 |
300 |